Pediatric Ventricular Assist Device (VAD) Guidelines
VAD Alarms and Malfunctions
Assessment of VAD alarms should begin with patient assessment. Evaluate the patient to identify
whether the alarm condition has created (or might have been caused by) changes in patient arousal,
hemodynamics, etc… Understanding whether and how the patient’s clinical condition has changed is
critical in evaluating the potential causes for alarms, especially alarms related to VAD flows. For VADs,
the following represents only a portion of the most common alarms, for a full list of alarms and
appropriate responses, refer to section on individual devices and the VAD folder on the Children’s
SharePoint Drive.
Low-flow alarms are the most common alarms encountered. On all devices, these may result from any
of the following circumstances:
1. Decreased preload
a. Hypovolemia
b. Right ventricular failure (during LVAD support)
c. Pulmonary arterial hypertension (during LVAD support)
2. Pericardial tamponade
3. Increased afterload
Treatment will depend on underlying cause, clinical evaluation is required, including echocardiography
as needed to assess preload and right ventricular function and to rule-out tamponade.